Reference Manual IQ Express Docking Station for ToxiPro, ToxiLtd & Toxi3Ltd Honeywell Analytics 651 South Main Street Midd

Seite 2

9“Next”. To install the files to another directory, click “Browse” and use Windows Explorer to specify the new location. Click “Next” once this is

Seite 3 - Table of Contents

10whether it is a powered or non-powered hub. Once the power requirement for each dock is satisfied the USB connection to the PC can be made by usin

Seite 4 - Warnings and Cautions

11capacity of 3 liters per minute, so it can be used to calibrate up to 4 docks simultaneously. Do not connect more than 4 docks to a single cylind

Seite 5

12 Air Filter at Dock Check Valve(s) Single Dock in Fresh Air Yes N/A Multiple Docks in Fresh Air Yes N/A Single Dock w/Zero Air Cylinder No No M

Seite 6 - 1. Overview

13when a cylinder is used to provide fresh air to a single dock. The fresh air filter is also not required in this configuration. For multiple dock

Seite 7 - Connection

143. Connect the pieces of tubing above the check valves to filter that was installed on the dock in step 2. At this point, the tubing assembly sh

Seite 8 - 2. Installation

15Once the IQ Express Software is launched, the software checks the dock for any records stored in dock’s memory and downloads them to the databas

Seite 9

16If a fault is detected and the IQ Express Dock is unable to complete the tests, the red fault light will be lit on the screen. Details of the fau

Seite 10 - 2.6 Power and Connectivity

17To darken or lighten the LCD on the IQ Express Dock, move the slider as appropriate. Battery The battery level shown on the dock tab is for the in

Seite 11 - Dock(s)

18 To change the gas type, click on the down arrow to the right of the gas setting and select the new gas type. Once the gas type is chosen, enter

Seite 12

1 THE IQ EXPRESS IS DESIGNED TO INTERFACE WITH HONEYWELL ANALYTICS GAS DETECTORS. HONEYWELL ANALYTICS GAS DETECTORS HAVE BEEN DESIGNED FOR THE DE

Seite 13 - Contaminated Environment

19With the O2 Bump and Calibration option enabled, when an oxygen detector is placed in the dock, the system will proceed with an oxygen bump test b

Seite 14 - Manifold

20 The output box is updated as the instrument is processed. If any faults occur, the details of the fault will be shown and a red fault light will

Seite 15 - 3. Software

21 7. Enter a unique 4-digit PIN in the input box under Access PIN and click on “Write to Instrument”. 8. Click “Set Config” to tell the dock to u

Seite 16 - 3.1 Dock Status

22 Note: Instrument firmware version 4.3 or greater is required for interface with the IQ Express. Attempts to use the IQ Express with ToxiPro or

Seite 17 - Instrument Status

23If the same instrument is placed in the IQ Express Dock immediately following its removal from the dock, the dock will show “ZERO Stabilizing” wit

Seite 18 - Gas Configuration

24 Instrument Settings for the detector including the Timeout Setting and the Calibration Due Reminder Setting  Settings for Printing Calibratio

Seite 19 - Special Instructions: Gas

25on your local hard drive under Program Files. New firmware files for ToxiPro must be stored in: C:\ Program Files \ Biosystems \ Flash Upload U

Seite 20 - 3.2.2 Instrument Tab

26 Click Yes. The status of the upload will be shown on he dock’s information page. The dock will also indicate that the Flash is being uploaded.

Seite 21 - Passcode

27Appendix A: USB Architecture with Network Access

Seite 22 - 4. Dock Use

28Appendix B: Calibration Frequency One of the most common questions that we are asked at Honeywell Analytics is: “How often should I calibrate my

Seite 23 - 4.1 Test Sequence

2Table of Contents Warnings and Cautions... 3 A. Signal Words.

Seite 24 - 5. Tools

29Honeywell Analytics Warranty Gas Detection Products General Honeywell Analytics, Inc. (hereafter Honeywell) warrants gas detectors, sensors and ac

Seite 25 - 5.2 ToxiPro

3AC Power... 16 LCD Contrast Control

Seite 26 - 7. Software Revisions

4CAUTION used without the safety alert symbol indicates a potentially hazardous situation which, if not avoided, may result in property damage.

Seite 27

51. Overview The IQ Express is an automatic calibration station for use with the ToxiPro, and ToxiLtd Gas Detectors. The IQ Express automatically p

Seite 28

61.2.2 Ethernet Connection and E-mail IQ Express Docks with the Ethernet upgrade may be connected to the PC via USB or via a network. To connect v

Seite 29 - Frequency

7instruments for SO2, Cl2, NO2, NH3, HCN and PH3. 1.7 Default Calibration Gas Settings and Dock Compatibility Dock Setting Default Cal Gas Setting

Seite 30 - Sensor Warranty Periods

82.2 IQ Administrator Pro and PostgreSQL Database Server Installation The IQ Administrator Pro Program is contained on the IQ Express Installation